Description
Thermo Scientific™ Orion™ AQUAMate UV-Vis Spectrophotometer with North America, European and UK Power Cord, Orion, AQ8100
Features:
- Cost-effective solution for accurate, visible-only measurements
- Ideal for most routine concentration measurements
- Unsurpassed data quality throughout the entire UV to the near-IR regions of the spectrum
- Intuitive user interface
- Dual-beam optical geometry provides accurate measurements
- high intensity, instant-on xenon lamp has a five-year lifespan
- Despite its relatively small footprint and lightweight design, it boasts numerous features at a competitive price
- Easy-to-use glove-friendly touchscreen user interface
- Includes over 260 preprogrammed methods for common water and wastewater chemistries
- Performance verification tests ensure wavelength accuracy and instrument function
- User-customizable absorbance, % transmittance, and concentration testing methods
- One-point calibration ensures the highest accuracy from preprogrammed methods
- Superior signal-to-noise performance and fast wavelength scanning allows quick and reliable acquistion of high quality spectral data
- Accessible from top, front or side
- Able to accommodate cells up to 100 mm cells
- Optional accessory for test tubes up to 25 mm diameter and 150 mm height
- Removable, washable sample compartment liner with magnetic placement and hold-down
Included:
- Test Tube Holder for vials and test tube from 12 to 25 mm OD
- 10 mm square single-cell holder
- AquaMate Water Analysis Methods Software
- Dust Cover
- External Power Supply
- North America power cable (NEMA 5-15)
- European power cable (CEE 7/7)
- UK Power cable (BS 1362)
Applications:
- Drinking Water
- Wastewater
- Environmental
Specifications:
|
Baseline Flatness |
±0.002 A |
|
Optical Design |
Dual beam |
|
Spectral Bandwidth |
2 nm |
|
Light Source (Typical Lifetime) |
Xenon flash lamp (5 years) |
|
Detector |
Dual silicon photodiodes |
|
Wavelength Repeatability |
±0.2 nm |
|
Wavelength Scanning Speed |
Slow/Medium/Fast - up to 1600 nm/min. |
|
Wavelength Data Resolution |
0.2, 0.5, 1, 2, and 5 nm |
|
Photometric Measurement Modes |
Absorbance, % transmittance, concentration |
|
Photometric Repeatability1 |
±0.001 A at 1 A measured at 1.0 A at 546 nm |
|
Photometric Noise2 |
≤0.00020A at 0 A at 260 and 500 nm ≤0.00030A at 1 A at 260 and 500 nm ≤0.00040A at 2 A at 260 and 500 nm |
|
Photometric Drift3 |
<0.0005 A/Hr (at 500 nm after warm-up) |
|
Photometric Stray Light |
< 1.0% T 198 nm (KCl) , <0.05% T at 220 nm (NaI), <0.03% T at 340 nm (NaNO2) |
|
Display |
7-inch high-definition color touchscreen, 800 × 1280 pixels |
|
Keypad |
Glove compatible touchscreen |
|
Printer Included |
Snap-on printer included |
1Measured at 1 A at 546 nm
2RMS at 500 nm (60 consecutive measurements)
3At 500 nm after 1-hour warm-up
Specifications
| Range | Wavelength: 190 to 1100 nm |
|---|---|
| Range | Photometeric: -2 A to 3.5 A; >1000 % T; to 0% T; 9999 C |
| Power | 100-240 volts, 50-60 Hz (auto-selected) |
| Accuracy | Wavelength: +/-0.5 nm, Photometric: ±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A |
| Compliant/Standard | CE / TUV |
| Connectivity | Single USB-A supports flash memory devices for method and data storage, Duplex USB-A supports connection to a Windows co |
| Connectivity | Export data to network or PC via Ethernet or Wi-Fi, Print via USB, ethernet, or Wi-Fi |
| Dimensions | 35.5 × 38.5 × 19.5 cm (L × W × H) |
| Weight | 7.5kg |
| Manufacturer | Thermo Scientific Orion |
| Model No | AQ8100 |
| Warranty | 1 year |