Features:

  • A manually reset device set to detect external chamber temperatures of between 220 and 240 degrees C
  • Final fail-safe device in protecting the autoclave chamber from an overheating condition
  • Consists of a contact box located next to the incoming power at the back of the maching and a sensing probe located in the upper channel of the rearmost heating element
  • Sensing probe and tube leading back to contabox are filled with liquid and should not be punctured or kinked so that the device operates properly
  • If the cut-out thermostat detects an over temperature and is activated, all power to the unit will be turned off and can only be reset once the unit has cooled down and the red reset button is pressed on the back of the machine