Humidity Test Chamber, 24.7 Cu Ft (700 L), 200 - 240 V (50/60 Hz), Binder Inc, 9020-0325

Item: 91-2677-26-EA

Description

Binder® Humidity Test Chamber, 24.7 Cu Ft (700 L), 200 - 240 V (50/60 Hz), 9020-0325

Features:

  • Safe thanks to climatic homogeneity that far exceeds accuracy required by ICH guidelines, even with fully loaded unit
  • Reliable thanks to fail-safe operation without compromise
  • Iinterior and long-life evaporator plate made entirely from stainless steel
  • Smart – wide range of accessories makes it highly compatible for adaptation to specific customer requirements
  • Economical thanks to maximum effective volume which allows for 30% larger load compared to competition
  • Temperature range: 0 °C to +70 °C
  • Humidity range: 10 % to 80 % RH
  • APT.line™ preheating chamber technology
  • Humidity regulation with capacitive humidity sensor and vapor humidification
  • Intuitive touchscreen controller with time-segment and real-time programming
  • BINDER Multi Management Software APT-COM™ Basic Edition
  • Internal data logger, measured values can be read out in open format via USB
  • Unit self-test for comprehensive status analysis
  • Tight-sealing inner door made of safety glass (ESG)
  • Avoidance of glass corrosion by special TIMELESS coating
  • Inner chamber made of stainless steel
  • 2 stainless steel racks
  • Access port with silicone plug, 30 mm, left
  • 4 stable castors, two with brakes, from 240 liters
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
  • Computer interface: Ethernet
  • Door heating

Specifications:

Option model Standard
Order number 9020-0325
Max. heat compensation at 40 °C [W] 600
Temperature fluctuation [± K] 0.1
Temperature range [°C] 0…70
Temperature variation at 40 °C [± K] 0.2
Humidity fluctuation at 25 °C and 60 % RH 1,5 ± % RH
Humidity fluctuation at 40 °C and 75 % RH 1,5 ± % RH
Humidity range 10…80
Recovery time after 30 seconds door open at 25 °C and 60 % RH [min] 16
Recovery time after 30 seconds door open at 40 °C and 75 % RH [min] 17
Temperature fluctuation at 25 °C and 60 % RH [± K] 0.1
Temperature fluctuation at 40 °C and 75 % RH [± K] 0.1
Temperature range [°C] 10…70
Temperature variation at 25 °C and 60 % RH [± K] 0.2
Temperature variation at 40 °C and 75 % RH [± K] 0.2
Nominal power [kW] 3.1
Phase (Nominal voltage) 1~
Power frequency [Hz] 50/60
Rated Voltage [V] 200…240
Unit fuse [A] 16
Inner doors 2
Unit doors 2
Depth net [in] 35.07
Height net [in] 75.8
Width net [in] 49.3
Depth [in] 22.69
Height [in] 49.3
Width [in] 38.34
Interior volume [cu.ft.] 24.7
Load per rack [lbs] 99
Net weight of the unit (empty) [lbs] 688
Permitted load [lbs] 331
Wall clearance back [in] 3.94
Wall clearance sidewise [in] 11.82
Sound-pressure level [dB(A)] 53
Energy consumption at 40 °C and 75 % RH [Wh/h] 620
Number of shelves (std./max.) 2/15

Specifications

Specifications
Range Humidity Range: 10% to 80% RH
Range Temperature Range: 0 degrees to 70 degrees C
Power 200 - 240V, 1 PH, 50/60 Hz
Dimensions 49.3 in. W x 75.8 in. H x 35.07 in. D
Weight 681 lbs.
Manufacturer Binder
Model No 9020-0325

You may also like