pH Meter, Waterproof, Extended Range Portable with mV, ORP and Temperature, Probe, Solutions, Case, Hanna Instruments, HI991003

Item: 21-2665-11-EA

UOM: EA (1)
Price:
Sale price$679.99

Description

 Features:

  • pH Sensor Check™; Allows users to check electrode status at any time
  • Multi-level LCD display; Display the current measurement simultaneously with the current temperature.
  • On-screen tutorial messages
  • Automatic Temperature Compensation
  • Automatic one or two point calibration
  • HI 1297 pH/ORP/Temperature and HI 1296 pH/Temperature probe in titanium enclosure
  • BEPS; (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
  • Battery % displayed on startup.
  • Compact, heavy-duty, and waterproof

Only one probe is required with this meter to measure pH, pH-mV (to check the electrode status), ORP and temperature, simultaneously. With all other instruments you need to switch the pH electrode with an ORP electrode (or vise-versa). The compact instrument fits in the palm of your hand and the bottom probe connections ensures the electrode cable does not get in your way. The unique Sensor Check™ feature allows the user to establish the status of the pH electrode before performing measurements. The Sensor Check™ feature consists of a pH-mV scale that shows the characteristics (slope and offset) of the sensor, allowing the user to determine if the electrode is in good condition.

This pH meter is protected in a waterproof casing for use in humid and wet environments. It also features a large dual-level LCD that displays pH or ORP and temperature readings simultaneously.The included pH/ORP temperature probe features an easy to clean recessed tip that prevents solids in solutions from collecting on the sensor. The AISI 316 stainless steel and titanium body of these probes function as a potential matching pin for increased stability of readings and extended sensor life.

Applications:

  • Plating baths
  • Wastewater
  • Swimming pools and spas
  • Environmental
  • Potable Water sanitizing processes

Specifications:

Range pH -2.00 to 16.00 pH
pH-mV ±825 mV (pH-mV)
ORP ±1999 mV
Temperature -5.0 to 105.0°C; 23.0 to 221.0°F
Resolution pH 0.01 pH
pH-mV 1 mV
ORP 1 mV
Temperature 0.1°C/0.1°F
Accuracy@ 20°C pH ±0.02 pH
pH-mV ±1 mV
ORP ±2 mV
Temperature ±0.5°C up to 60°C; ±1.0°C (outside); ±1.0°F up to 140°F; ±2.0°F (outside)
pH Calibration  automatic one or two point calibration with two sets of memorized buffers (standard 4.01, 7.01, 10.01 or NIST 4.01, 6.86, 9.18)
TemperatureCompensation  automatic, -5.0 to 105.0°C (23.0 to 221.0°F)
Electrode  HI 1297D pre-amplified pH/ORP probe with internal temperature sensor, DIN connector and 1 m (3.3’) cable (included)
BatteryType/Life  1.5V (3) AAA / approximately 1200 hours of continuous use. auto-off after eight minutes of non-use
Environment  0 to 50°C (32 to 122°F); RH max. 100%
Dimensions  152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)
Weight  205 g (7.2 oz.)

 

Supplied with:

  • HI 991003 portable meter
  • HI 1297D pH/ORP probe with internal temperature sensor
  • HI 70004 pH 4.01 buffer sachet
  • HI 70007 pH 7.01 buffer sachet
  • HI 700661 electrode cleaning solution sachet (2)
  • 3 (1.5V) AA batteries
  • Rugged carrying case
  • Instructions Manual