Description
Natural Gas Analyzer, Gas Chromatograph, for Gas Power Plant Applications, Model 8610C, USB 2.0 Chassis, 115 Vac, 60 Hz
Features:
- TCD Detector
- 10-port Gas Sampling Valve
- 1-meter Hayesep-D Packed Column
- 30-meter, 0.53 mm MXT-1 Capillary Column
- 1 channel PeakSimple Data System
- ...on the compact 8610C chassis
The Natural Gas Analyzer GC system is pre-plumbed and ready to measure N2, methane, CO2 , ethane, H2 , O, propane, iso- and normal pentanes, and C6 plus backflush. The main benefits of the Natural Gas Analyzer GC are simplicity, low cost, and the ability to determine the water content of the gas. The Natural Gas Analyzer GC uses just a sigle 10-port gas sampling valve and two columns, and is tolerant of valve timing variations or operator adjustments.
Unlike widely used competitive micro GCs. the Natural Gas Analyzer GC is not only tolerant of water in the sample gas, but it actually generates a quantitative water peak. The seven minute analysis may be longer than a micro GC analysis, but the Natural Gas Analyzer GC does not need baking out between runs. Therefore, about the same number of runs may be made per day. with the Natural Gas Analyzer GC as with a typical micro GC.
Includes:
- Ambient-to-400°C temperature-programmable virtual column oven
- Single cold on-column direct injection port
- Electronic pressure control (EPC) of all GC system gases
- Pressure-programmable carrier gas
- Thermal Conductivity detector (TCD)
- Automated gas sampling valve mounted in a thermostatically-controlled valve oven
- One capillary and one packed analytical column
- LED "at-a-glace" display of all controlled zones, valve status
- Digital LED display of temperature, pressure and voltage
- IBM PC-compatible PeakSimple data system software for Widows XP (compatible with Windows Vista/7/8 environments)
- Built-in 1-ch USB 2.0-interfaced data acquition hardware,
- 6' USB 2.0 cable (A to B connectors)
- GC accessory kit
- Operation / service manual CD-ROM
- Two-Year Warranty